11th World Congress on Industrial Process Tomography
Multi-beam computed tomography scanner design
for industrial multi-phase flow investigations
A. Bieberle*,1, A. Döß1, M. Löschau2, E. Schleicher1, U. Hampel1,3
1 Helmholtz-Zentrum Dresden - Rossendorf, Bautzner Landstr. 400,
01328 Dresden, Germany
2 HZDR Innovation GmbH, Bautzner Landstraße 400,
01328 Dresden, Germany
3Chair of Imaging Techniques in Energy and Process Engineering,
Technische Universität Dresden, 01062 Dresden, Germany
*Email: a.bieberle@hzdr.de
ABSTRACT
We present a multi beam gamma ray computed tomography (CT) scanner design for multi-phase flow investigations in industrial apparatuses. It mainly comprises a collimated 137Cs isotopic source and an in-house developed detector arc with overall 16 scintillation detectors offering a quantum efficiency of approximately 75% and an active area of 10×10 mm2 each. The detectors are operated in pulse counting mode to enable gamma photon energy discrimination. A key element for highest flexible application of the proposed CT scanner is the elaborated detector design that is concerted with a sophisticated scanning procedure that allows for multi beam projection acquisition. Thus, scanning times can be adopted to various object sizes to save measuring times while the arrangement of source and detector arc remains identical.
Keywords: gamma ray computed tomography, detector technology
Industrial Application: Chemical industry and General
Sign-in to access the full text
Copyright © International Society for Industrial Process Tomography, 2023. All rights reserved.