Adaptive Signal Excitation Source Design for Wearable Multimodal Electrical Impedance Tomography System

Y. K. Chang1, S. J. Ren1*, C. Tan1, F. Dong1

1School of Electrical and Information Engineering, Tianjin University
Tianjin Key Laboratory of Process Measurement and Control Tianjin 300072, China

*Email: Rensjie@tju.edu.cn


ABSTRACT

Addressing the challenge of current output stability in the Wearable Multimodal Electrical Impedance Tomography (WMEIT) system under dynamic environments, this study proposes an adaptive signal excitation source design. This design is based on a Voltage-Controlled Current Source (VCCS) architecture, regulated through a digitally programmable potentiometer, and employs an incremental proportional control algorithm to provide real-time compensation for load variations within the range of 250 Ω–7.5 kΩ. Experimental results demonstrate that at a 20 kHz excitation frequency, the system output current remains stable at the set value "0.02 mAₚₚ (error <1%). Through frequency response optimization, 30 dB of interference attenuation is achieved in the ECG frequency band (100 Hz), effectively eliminating cross-talk between multimodal signals. This provides critical hardware support for dynamic cardiopulmonary monitoring under motion conditions (such as pulmonary tidal volume assessment and cardiac-gated imaging).


Keywords: Electrical Impedance Tomography, Wearable System, excitation source